How often to inspect an AED
Check it at least once a month. Also check it after every use, after it has been moved, and whenever a pads or battery date is coming up. A routine check takes a few minutes, and your AED maker’s instructions come first if they ask for more.
The 10-point AED inspection checklist
The first six are essential on every AED. The last four apply if your AED has them, and most sites do.
- Readiness indicator. Shows ready: a blinking green light, a green ✓ or a steady green light, depending on the model. What each light means.
- Serial number. The label is there and readable. Note the number in your log: it also tells you how old the AED is. Read its age.
- Pads expiry. The date on the pads package has not passed, and the package is sealed and undamaged. Where the date is.
- Battery expiry. The date on the battery label has not passed. What the label means.
- Battery fitted. The battery is pushed fully in, flush with the case, and latched where there is a latch.
- Pads connected. The pads are plugged into the AED (or the cartridge is fitted) so they are ready when the lid opens.
- Child pads or key. If the AED has them: present, in date, and kept beside the AED. They are for children under 8 or under 25 kg.
- Cabinet or case. Easy to reach, not blocked or locked shut, clearly signed, and its alarm works if it has one.
- Rescue kit. Gloves, a razor, scissors and a CPR face mask, sealed and kept with the AED.
- Emergency contacts sticker. On the AED or its cabinet and readable, with your local emergency number.
Each model hides its dates in different places. The guide for your model shows exactly where.
Printable AED inspection checklist
Print this and keep it with the AED, or use the free app and let the report be your record.
| Check | Pass? | Notes |
|---|---|---|
| Readiness indicator | ☐ Yes ☐ No | |
| Serial number | ☐ Yes ☐ No | |
| Pads expiry | ☐ Yes ☐ No | |
| Battery expiry | ☐ Yes ☐ No | |
| Battery fitted | ☐ Yes ☐ No | |
| Pads connected | ☐ Yes ☐ No | |
| Child pads or key | ☐ Yes ☐ No | |
| Cabinet or case | ☐ Yes ☐ No | |
| Rescue kit | ☐ Yes ☐ No | |
| Emergency contacts sticker | ☐ Yes ☐ No | |
| Date and inspected by | ||
What to do when a check fails
- Not showing ready: check that the battery is fully fitted and the pads are plugged in, then look again. If it still isn’t ready, contact your supplier, put a notice on the cabinet, and tell people where the nearest working AED is.
- Pads or battery past their date: replace them. Order a month or two before the date, so the new ones arrive in time.
- Missing rescue kit or sticker: replace them. They’re cheap, and a razor or scissors can be the difference in getting the pads on quickly.
Run the checklist on your phone
AED SmartX Inspector walks you through these ten checks with your phone’s camera. It films the readiness light, reads the serial number and every expiry date with AI, scores the AED out of 100, and emails you a PDF report. The quick check (readiness light and serial number) takes about a minute; the full inspection, about six. It’s free, in English and Hindi.