Is this your AED?
Lime green and upright, with a colour screen on the front. The model name is printed on the front of the unit and on the label at the back.
1. Is it ready? Check the readiness light
Every AED tests itself and shows the result on a readiness indicator. On the AED 3 it is the small status window just right of the On/Off button.
- Ready: A green ✓ (tick) in the window.
- Not ready: If the green ✓ isn’t showing, it needs attention.
See what the light means on other models in our status light guide.
2. Find the serial number
The serial number label on the back, just above the battery.
How old is your AED 3?
After its first letter “A”, the serial follows ZOLL’s pattern: “X”, two digits for the year, then a letter for the month (A is January, L is December). “AX19D…” was made in April 2019.
Age matters: many AED owners plan to replace a unit at around ten years, and an older unit may no longer be supported by its maker. More in how old is my AED.
3. Check the pads expiry date
On the pads package in the back of the AED, beside the ⌛ symbol.
Replace the CPR Uni-padz if the date has passed, after any use (they are single-use), or if the package is open or damaged. Expired pads can dry out and may not stick or conduct properly. More in AED pads expiry.
4. Check the battery date
The white label on the battery at the back shows its install-by date.
More in AED battery expiry.
5. Make sure the battery is fitted
The battery should be clicked in, flush with the case at the back.
6. Make sure the pads are connected
The pads cable should be plugged into its socket at the top right of the front.
Child pads or key
CPR Uni-padz are used for children as well as adults, so most AED 3 owners don’t need separate child pads. ZOLL’s Pedi-padz II are also available.
The cabinet, rescue kit and emergency sign
- Cabinet or case: easy to reach, not locked or blocked, and the alarm working if it has one.
- Rescue kit: gloves, a razor, scissors and a CPR face mask, sealed and with the AED.
- Emergency contacts sticker: on the AED or its cabinet, with your local emergency number.
If something fails
Many faults take seconds to fix: push the battery fully home, plug the pads back in, close the lid, then check the readiness light again. An expired battery or pads can’t be fixed on the spot and need replacing. The free inspection’s report lists exactly what failed, and you can ask for a quote for the right pads or battery for your AED 3 from the result screen.