ZOLL · inspection guide

How to inspect a ZOLL AED Plus

The monthly check, step by step: the readiness light, the pads and battery dates, and the serial number that tells you how old it is.

By the Think Health team, AED suppliers · Updated

ZOLL AED Plus AED

The AED Plus check in six steps

  1. Readiness light: a green ✓ (tick) in the window.
  2. Serial number: readable, and tells you how old the unit is.
  3. Pads expiry: the date has not passed.
  4. Battery date: the date has not passed.
  5. Battery fitted: pushed fully in and flush.
  6. Pads connected: plugged into the AED.

Is this your AED?

Bright green, with the handle moulded into the top. The model name is printed on the front of the unit and on the label at the back.

1. Is it ready? Check the readiness light

Every AED tests itself and shows the result on a readiness indicator. On the AED Plus it is the status window on the left of the handle.

  • Ready: A green ✓ (tick) in the window.
  • Not ready: A red ✗ means it is not ready: it failed a self-test, often because the batteries are low. It may also beep.

See what the light means on other models in our status light guide.

2. Find the serial number

The barcode label on the back, just below the handle.

How old is your AED Plus?

ZOLL puts the year and month in the serial number: after the “X” come two digits for the year, then a letter for the month (A is January, L is December). “X14K718292” was made in November 2014.

Age matters: many AED owners plan to replace a unit at around ten years, and an older unit may no longer be supported by its maker. More in how old is my AED.

3. Check the pads expiry date

On the pads pack or its box, beside the ⌛ symbol.

Replace the CPR-D-padz if the date has passed, after any use (they are single-use), or if the package is open or damaged. Expired pads can dry out and may not stick or conduct properly. More in AED pads expiry.

4. Check the battery date

The “Replace batteries on or before” label, just below the status window on the handle.

The AED Plus runs on ten lithium cells. Replace all ten together, never one at a time.

More in AED battery expiry.

5. Make sure the battery is fitted

The battery compartment cover on the back should be closed and latched, with all ten cells seated.

6. Make sure the pads are connected

Lift the cover and check the pads cable is plugged into its socket.

Child pads or key

ZOLL’s child pads are Pedi-padz II. Keep a spare pack if children could be present.

The cabinet, rescue kit and emergency sign

  • Cabinet or case: easy to reach, not locked or blocked, and the alarm working if it has one.
  • Rescue kit: gloves, a razor, scissors and a CPR face mask, sealed and with the AED.
  • Emergency contacts sticker: on the AED or its cabinet, with your local emergency number.

If something fails

Many faults take seconds to fix: push the battery fully home, plug the pads back in, close the lid, then check the readiness light again. An expired battery or pads can’t be fixed on the spot and need replacing. The free inspection’s report lists exactly what failed, and you can ask for a quote for the right pads or battery for your AED Plus from the result screen.

ZOLL AED Plus: common questions

What does the red X on a ZOLL AED Plus mean?

The AED is not ready to use: it failed its self-test. Low batteries are a common cause. Replace all ten batteries together and check that the window shows a green ✓. If it still shows a red ✗, contact your supplier.

How do I know my ZOLL AED Plus is ready?

Look at the status window on the left of the handle. A green ✓ means it passed its last self-test and is ready.

Where is the battery date on a ZOLL AED Plus?